| Current Transformer On-site Calibrator DARUI Electric recommends Main technical indicators: >Current transformer error measurement part ①Overall accuracy: 1/3 of the error limit of the tested current transformer ②Test range: 5A/5A~25000A/5A or 5A/1A~6300A/1A ③Working range of tested current transformer: 1% to 200% ④Secondary burden: 2.5VA~300VA, COSφ=0.1~1.0 ⑤Accuracy class of tested current transformer: 1.0, 0.5, 0.5S, 0.2 and 0.2S ⑥Resistance and admittance measurement error ≤5.0% >Measurement range: R: 0.00Ω~20.0Ω Y: 0.000mS~100.0mS >Power consumption: 20VA >Accuracy class: 0.05S >Maximum dimensions (cm): 460×375×183 >Weight (kg): 10 Current Transformer On-site Calibrator has a built-in standard voltage transformer with the same ratio as the tested current transformer, with accuracy up to 0.05 class, accurately measuring the ratio and no-load error of the tested current transformer. Then, combined with the test results of impedance and admittance, the error of the transformer is calculated. >Uses a near-power-frequency different-frequency power source for testing to prevent on-site power-frequency electromagnetic radiation and series interference. >Wide measurement range, up to 5A/5A~25000A/5A or 5A/1A~6300A/1A. >Has the function of testing the ratio and secondary winding internal resistance of current transformers. >Uses an 800×600 high-resolution large-screen color LCD display, with user-friendly interface and operation design, and touch screen auxiliary operation, making operation more convenient and faster. >Uses precise software algorithms to further improve the accuracy of measurement data. >Intelligently judges external wiring conditions, prompts wiring errors, ratio errors, polarity errors, etc. >Automatically rounds test data and judges whether it exceeds the tolerance. Out-of-tolerance data is displayed in orange, and the bottom right corner of the window shows out-of-tolerance, making the data characteristics of the transformer intuitive and clear. >Directly issues on-site verification conclusions, qualified or out-of-tolerance. >Large-scale memory can store up to 1000 sets of on-site test data. >Uses RS232 or USB interface to connect to a computer to print data certificates. >Uses engineering plastic mold casing, shockproof and pressure-resistant, ensuring the safety of on-site operators and equipment. Current Transformer On-site Calibrator also has the functions of measuring current transformer errors by the recursion method and voltage transformer errors by the potential difference method, facilitating on-site verification of metering devices. >On-site verification of current transformers does not require standard current transformers, current boosters, burden boxes, voltage regulators, or high-current cables. It uses extremely simple test wiring and operation to achieve verification of current transformers, greatly reducing work intensity and improving work efficiency, facilitating on-site verification of transformers. Note: The above information and technical parameters are compiled and released by Yangzhou Darui Electric Co., Ltd. |