Anti-interference Dielectric Loss Tester
Performance Characteristics
The automatic anti-interference precision dielectric loss tester is used for on-site anti-interference dielectric loss measurement, or for precise dielectric loss measurement in the laboratory. The instrument is an integrated structure, with built-in dielectric loss bridge, variable frequency power supply, test transformer, and standard capacitor, etc. It adopts variable frequency anti-interference and Fourier transform digital filtering technology, fully automatic intelligent measurement, and the measurement data is very stable under strong interference. The measurement results are displayed on a large screen LCD, and a built-in micro printer can print out the results.
Technical Specifications
1. Dielectric loss and capacitance measurement
Accuracy: Cx: ±(reading×1%+1pF)
tgδ: ±(reading×1%+0.00040)
Anti-interference index: Variable frequency anti-interference, the above accuracy can still be achieved under 200% interference
Capacitance range: Internal high voltage: 3pF~60000pF/10kV 60pF~1.2μF/0.5kV
External high voltage: 3pF~1.5μF/10kV 60pF~30μF/0.5kV
Resolution: Up to 0.001pF, 4 significant digits
tgδ range: unlimited, resolution 0.001%, automatic identification of three types of test objects: capacitance, inductance, and resistance.
Test current range: 10μA~5A
Internal high voltage: Set voltage range: 0.5~10kV
Maximum output current: 200mA
Voltage regulation method: continuous smooth adjustment
Voltage accuracy: ±(1.5%×reading+10V)
Voltage resolution: 1V
Test frequency: 45~65Hz integer frequency
49/51Hz, 45/55Hz automatic dual frequency conversion
Frequency accuracy: ±0.01Hz
External high voltage: Maximum test current 5A/40~70Hz for normal wiring
Maximum test current 10kV/5A/40~70Hz for reverse wiring
CVT self-excitation method low voltage output: output voltage 3~50V, output current 3~30A
2. CVT ratio measurement
Ratio range: 10~99999
Ratio accuracy: ±reading×1%
Phase range: 0~359.999°
Phase accuracy: ±0.02°
3. Input power: 180V~270VAC, 50Hz/60Hz±1%, mains or generator power supply
4. Computer interface: standard RS232 interface
5. Printer: built-in micro thermal printer
6. Ambient temperature: -10℃~50℃
7. Relative humidity: <90%, no condensation